Skip to main content

Table 3 Accuracy, precision, LOD and LOQ data of analytical procedure

From: A revised method for determination of dialkylphosphate levels in human urine by solid-phase extraction and liquid chromatography with tandem mass spectrometry: application to human urine samples from Japanese children

 

Concentration (μg/L)

n

DMP

DMTP

DEP

DETP

Within-run precision (%RSD)

1

4

  

10.3

6.4

2

4

3.1

9.8

10.7

3.4

5

4

3.1

2.7

3.9

4.5

10

4

3.6

5.8

  

20

4

  

5.5

3.2

100

4

6.7

3.6

  

Mean recoverya

1

2

95

64

81

101

5

2

100

105

82

93

Between-run precision (%RSD)

1

7

6.9

7.7

9.6

7.6

10

7

9.0

4.3

7.5

7.8

R 2 of calibration line

  

0.999

0.998

0.997

0.999

LOD (μg/L) (S/N = 3)

  

0.4

0.4

0.2

0.1

LOQ (μg/L) (S/N = 10)

  

1.2

1.2

0.6

0.3

  1. n number of observations, RSD relative standard deviation, LOD limit of detection, LOQ limit of quantitation
  2. aRecovery obtained by adding the standards just before the separation analysis step