Skip to main content

Table 2 Compound-specific mass spectrometer settings

From: A revised method for determination of dialkylphosphate levels in human urine by solid-phase extraction and liquid chromatography with tandem mass spectrometry: application to human urine samples from Japanese children

Compounds

Fragementor (V)

Precursor ion (m/z)

Product ion (m/z)

Collision energy (eV)

Polarity

Retention time (min)

DMP-d6

90

133

115

18

Positive

6.48

DMP

90

127

109 (Q)

95 (C)

20

20

Positive

6.48

DMTP-d6

90

147

129

14

Negative

6.54

DMTP

90

141

126 (Q)a

96 (C)

13

21

Negative

6.54

DEP-d10

80

165

101

16

Positive

6.53

DEP

80

155

127 (Q)

99 (C)

10

10

Positive

6.53

DETP-d10

100

179

95

20

Negative

6.89

DETP

90

169

95 (Q)

141 (C)

17

9

Negative

6.89

  1. C confirmation ion, Q quantification ion
  2. aTotal of Q and C were used for DMTP quantification due to its low sensitivity